Difference between revisions of "XPS50P50"
From CPS Technical Wiki
(→Wiring diagrams) |
(→Power input connector) |
||
Line 3: | Line 3: | ||
==Connector pinouts== | ==Connector pinouts== | ||
− | === | + | ===Monitor connector=== |
+ | This connector provides access to metering output signals and the safety interlock input signal. | ||
+ | |||
+ | Pin Name Function Usage | ||
+ | --- ------ -------- ----- | ||
+ | 1 RETURN Signal ground Ground reference for meter outputs (kV MON, mA MON). | ||
+ | 2 kV MON HV meter (out) 0 to +10 V = 0 to +50 kV out. | ||
+ | 3 mA MON Beam meter (out) 0 to +10 V = 0 to 1 mA out. | ||
+ | 4 INTLK Safety interlock (in) Short to pin 1 to enable HV output; leave open to disable HV output. | ||
===Control/monitor connector=== | ===Control/monitor connector=== |
Revision as of 16:05, 7 December 2021
This is the wiki page for CPS model XPS50P50.
Contents |
Connector pinouts
Monitor connector
This connector provides access to metering output signals and the safety interlock input signal.
Pin Name Function Usage --- ------ -------- ----- 1 RETURN Signal ground Ground reference for meter outputs (kV MON, mA MON). 2 kV MON HV meter (out) 0 to +10 V = 0 to +50 kV out. 3 mA MON Beam meter (out) 0 to +10 V = 0 to 1 mA out. 4 INTLK Safety interlock (in) Short to pin 1 to enable HV output; leave open to disable HV output.
Control/monitor connector
Wiring diagrams
Model XPS50P50 can either be controlled locally, by adjusting its internal trimpots, or remotely by connecting it to external programming voltages.