This is the wiki page for CPS model XPS50P50.
 
This is the wiki page for CPS model XPS50P50.
 +
 +
==Connector pinouts==
 +
 +
===Monitor (J1)===
 +
 +
J1 provides access to metering output signals and the safety interlock input signal.
 +
 +
{| class="wikitable"
 +
|-
 +
! Pin !! Name !! Function !! Usage
 +
|-
 +
| 1|| RETURN|| Signal ground|| Ground reference for meter outputs (kV MON, mA MON)
 +
|-
 +
| 2|| kV MON|| HV meter (out)|| 0 to +10 V = 0 to +50 kV out
 +
|-
 +
| 3|| mA MON|| Beam meter (out)|| 0 to +10 V = 0 to 1 mA out
 +
|-
 +
| 4|| INTLK|| Safety interlock (in)|| Short to pin 1 to enable HV output; leave open to disable HV output
 +
|}
 +
 +
===Control (J2)===
 +
 +
J2 provides access to power supply control signals.
 +
 +
{| class="wikitable"
 +
|-
 +
! Pin !! Name !! Function !! Usage
 +
|-
 +
| 1|| VREF|| 10V reference (out)|| Reference voltage for external circuitry; 10 mA maximum.
 +
|-
 +
| 2|| NC|| No connect||
 +
|-
 +
| 3|| ISET|| Beam setpoint (in)|| Beam control voltage: 0 to +10V = 0 to 1 mA out. Local control: connect to pin 4. Remote control: connect to external 0 to +10V source.
 +
|-
 +
| 4|| IPOT|| Beam control pot (out)|| Local control: connect to pin 3. Remote control: leave disconnected.
 +
|-
 +
| 5|| NC|| No connect||
 +
|-
 +
| 6|| VSET|| HV setpoint (in)|| HV control voltage: 0 to +10 V = 0 to +50 kV out. Local control: connect to pin 7. Remote control: connect to external 0 to +10V source.
 +
|-
 +
| 7|| VPOT|| HV control pot (out)|| Local control: connect to pin 6. Remote control: leave disconnected.
 +
|-
 +
| 8|| RETURN|| Signal ground|| Ground reference for VREF, VPGM and IPGM.
 +
|-
 +
| 9|| RETURN|| Signal ground|| Ground reference for VREF, VPGM and IPGM.
 +
|}
  
 
==Wiring diagrams==
 
==Wiring diagrams==
  
[[File:Xps50 wiring.png|thumb|Wiring diagrams showing local vs. remote control of model [[XPS50P50]]]]
+
Model XPS50P50 can either be controlled locally, by adjusting its internal trimpots, or remotely by connecting it to external programming voltages. In both cases, the output voltage and beam current may be externally monitored by connecting meters (or analog-to-digital converters) to J1 pins 2 and 3.
 +
 
 +
[[File:Xps50 wiring.png|thumb|center|550px|Wiring diagrams showing local vs. remote control of model XPS50P50]]
  
  

Latest revision as of 16:27, 7 December 2021

This is the wiki page for CPS model XPS50P50.

Contents

Connector pinouts

Monitor (J1)

J1 provides access to metering output signals and the safety interlock input signal.

Pin Name Function Usage
1 RETURN Signal ground Ground reference for meter outputs (kV MON, mA MON)
2 kV MON HV meter (out) 0 to +10 V = 0 to +50 kV out
3 mA MON Beam meter (out) 0 to +10 V = 0 to 1 mA out
4 INTLK Safety interlock (in) Short to pin 1 to enable HV output; leave open to disable HV output

Control (J2)

J2 provides access to power supply control signals.

Pin Name Function Usage
1 VREF 10V reference (out) Reference voltage for external circuitry; 10 mA maximum.
2 NC No connect
3 ISET Beam setpoint (in) Beam control voltage: 0 to +10V = 0 to 1 mA out. Local control: connect to pin 4. Remote control: connect to external 0 to +10V source.
4 IPOT Beam control pot (out) Local control: connect to pin 3. Remote control: leave disconnected.
5 NC No connect
6 VSET HV setpoint (in) HV control voltage: 0 to +10 V = 0 to +50 kV out. Local control: connect to pin 7. Remote control: connect to external 0 to +10V source.
7 VPOT HV control pot (out) Local control: connect to pin 6. Remote control: leave disconnected.
8 RETURN Signal ground Ground reference for VREF, VPGM and IPGM.
9 RETURN Signal ground Ground reference for VREF, VPGM and IPGM.

Wiring diagrams

Model XPS50P50 can either be controlled locally, by adjusting its internal trimpots, or remotely by connecting it to external programming voltages. In both cases, the output voltage and beam current may be externally monitored by connecting meters (or analog-to-digital converters) to J1 pins 2 and 3.

Wiring diagrams showing local vs. remote control of model XPS50P50